The SYCON P is a titration device for measuring the +p value. In addition to the usual relays and input contacts, the online analyzer has a 20 mA current and RS485 interface with Modbus RTU protocol.
Monitoring of a freely programmable limit value
Measurement of the +p value
Graphic display, menu navigation in German, English, Italian, French, Spanish, Russian, Polish, Dutch and Danish.
Simple commissioning using the installation wizard
Documentation of all measurement results on an SD card
Low maintenance thanks to peristaltic pump technology and flushing the measuring cell with line pressure
Self-calibrating against ambient light, contamination and turbidity of the water sample
Reliable thanks to self-diagnosis and plausibility monitoring of sensor data during analysis
Maintenance-free full color sensor
Connections:
4 potential-free outputs:
Limit value exceeded
Limit value exceeded / limit value not reached / indicator level low
Device fault
Analysis active / cooling water valve / indicator level low
Current interface configurable as 0 – 20mA or 4 -20mA for measured value output or status message
Input for potential-free contact for external analysis triggering, water meter for volume interval or pausing the internal analysis interval
Input for remote rest via external control
Modbus RTU for querying the measured values and device parameters via a PLC